PTC-8100D Single Phase Meter Calibration Stand 
Operation manual
1. Introduction
PTC-8100D Single Phase Meter Calibration Stand is a new generation, high stability single phase energy meter calibration phantom load power source. It uses the advanced technology and electronic components available in 90 decade . The instrument can generate sine-wave voltage/current and phantom power to drive single phase energy meter, usually it is the main part of energy meter test bench.
1.1 Feature and function
1) Equipped with electronic phantom load power source, assuring high stability, security and large output capacity.
2) Automatic adjustment of current measurement range. Fault self-detection in voltage amplifier and current amplifier.
3) Equipped with RS-232C interface, can be operated from software.
2. Structure and main parts
① HMI Interf        ②  Keyboard           
③ Power switch      ④  U&I Output Terminal 
⑤ Pulse input port  ⑥  RS485 communication port(for meter)
3. Technical specification 
Specification of phantom load power source:
Power and consumption:
 Power supply:  1P2W  220V±10%,  50Hz / 60Hz;
Max consumption:  400VA 
Dimension:  480mm X 200mm X 450mm                                                      
Current source: 
Output current:  1×0.01A - 1×100A 
Adjusting resolution:  0.01% (min. 0.001A)
Power factor:  -1.0 ~ 1.0(0~3600;  resolution: 0.10)
Output frequency:  45Hz~65Hz;
THD(sine wave output):  <0.8%
Stability of current during sine wave output: 
0.05% (180s)
Output wave:  sine wave
Output capacity:   30VA per meter position                                                                                                                      

Voltage source: 
Output voltage:   1×30V - 1×300V 
Load:   The Load could be resistor / inductive / capacitive(<4uF)
Adjusting resolution:   0.01% (min. 0.1V)
Output frequency:   45Hz~65Hz;
THD (sine wave output):   <0.8%
Stability of voltage during sine wave output: 
0.05% (180s)
Output wave:  sine wave
Output capacity:   15VA per meter position

4.Goods list of one set equipment 
1) Single phase phantom load power source               1
2) Operation manual                                   1
5. Working principle and frame
Introduction: 
1)Signal generator
The   marked parts is included in signal generator board. Signal generator is equipped with DDS waveform technology (digital frequency modulation, amplitude modulation and phase modulation). Embedded MCU is the brain of signal generator, also it is the brain of the instrument, signal generator generate high stability sine waveform which is send to voltage amplifier and current amplifier.
2) Power amplifier
Because the instrument uses integrated PWM amplifier technology, the power amplifier features high stability / high reliability and high efficiency. The amplifier has well-designed protection, when output voltage shortcut or output current circuit open, the alarm system will auto shut down the amplifier, the same time LCD will display which part is out of order. 
3) Measurement
Error of tested meter is achieved by comparison with the reference meter.
6. Embed software user manual 
6.1  Definition of Keyboard and Symbol of Display Unit
Key define:
F4:  StartingUP test
F5:  No load test
5% ~ 600% ,Imax,0.5Imax:  Load select 
Start:  Output U&I
0 ~9:  Number
0.5L, 0.8L, 1.0, 0.8C, 0.5C:  Power Factor
.  :  Decimal
↑↓:  up/down or Increase/decrease current/voltage
    :  Enter key
Set:  tested meter’s parameter / System parameter
ESC: Back to previous menu
Adj: Adjust output voltage & output current
6.2  Initial screen
Turn on power for equipment, the first screen will display as below:
6.3  Main Screen
After system initial finished , the main screen will display as below:


① Tested meter’s model   ② Output or Stop indication
③ Error information
 [BZBErr: communication error with reference meter
 800Err: communication error with error calculator]
 ④ Output value   
 ⑤ Tested meter’s parameter and output parameter
 ⑥ Output waveform      ⑦ Help information 
6.4  Tested Meter’s Parameters And System Parameters
6.4.1  Parameter menu :
Parameter setting menu is shown as below, select corresponding menu item to set parameters about tested meter or about system.
Use number key or ↑/↓ key to select menu, Press ESC Key will back to main screen.
6.4.2  Set Parameters of tested meter [Menu item 1]
Select menu ‘tested meter para…’, Then press key ‘SET’ or ‘Enter’ to set tested meter’s parameters, the screen will display as below:
Use ↑↓ key to select corresponding setting item, press number key or decimal point to proceed setting. Press ↑↓ or press ESC will confirm any input data, press ESC will also confirm any input data and return to previous menu. The effective frequency range should be 45.00Hz – 65.00Hz .  C is the meter’s constant tested meter.  N is the test round when output voltage is 100%Un and output current is 100%Ib, Nmax is the maximum test round and Nmin is minimum test round, Class means tested meter’s class.
6.4.3  Waveform Parameter [Menu item 2]
Select menu ‘WaveformSelect…’, Then press key ‘SET’ or ‘Enter’ to select output waveform(note: KP-P1200-A can output sine wave only), the screen will display as below:
6.4.4  Startingup test/no load test parameters [Menu item 3]
Select menu ‘StartingupCreep para…’, Then press key ‘SET’ or ‘Enter’ to set tested meter’s Startingup test/no load test parameters, the screen will display as below:
6.4.5  System parameters [Menu item 4]
Select menu ‘SystemPara…’, Then press key ‘SET’ or ‘Enter’ to set parameters about test instrument, these parameter are very important, in order to avoid these parameters being changed easily, user should input a password firstly, the screen will display as below:
Password of the equipment is “6003”. After input password, the screen will display as below:
System parameter : System parameter include meter position / display selection and other system parameter. 
Output Parameter : Output Parameter include voltage output parameter and current output parameter. 
6.4.5.1  System parameter
System parameter include meter position / display selection and other system parameter. Screen will display as below:
Meter positions : meter positions of the test equipment.
Find mark function: when processing starting_up test or no_load test ,if Find mark function is enabled, before processing starting_up test or no_load test, software will process Find mark function firstly.
Disp Selection: user can select information displayed on the LCD, if select ‘Power’, the real output power will displayed on lcd when test equipment output voltage and current. if select ‘Phase’, the real output angle between voltage and current will displayed on lcd when test equipment output voltage and current.
Auto adjust : when testbench output voltage and current, embedded software will read real output value and real output angle between voltage and current, if the real output value or real output angle is different with the set value and if this function is enabled , embedded software will rectify the output value.
Language : select which language is used of the test equipment.
6.4.5.2 Output Parameter
Output Parameter : Output Parameter include voltage output parameter and current output parameter, screen will display as below:
Ua,Ub and Uc are amplitude coefficient of voltage output.
   [for KP-P1200, only Ua parameter is use ]
Ia, Ib and Ic are amplitude coefficient of current output.
   [for KP-P1200, only Ia parameter is use ]
ΦIa, ΦIb andΦIc is phase coefficient.
   [for KP-P1200, only ΦIa  parameter is use ]
ΦUb andΦUc is Ub and Uc phase coefficient.
Iqa, Iqb and Iqc is offset coefficient of current output.
(When coefficient being changed, the output value will be changed )
6.5  F1 Function (Starting up test)
Press F1 key software will begin ‘Starting up test’, LCD screen display as below
(  If ‘Find mark function’ is enabled(in system parameter)  ):
If ‘find mark function’ is enabled(in system parameter), embedded software will process ‘Find mark’ function firstly, if ‘find mark’ function is disabled, this screen won’t display.
If ‘Find mark’ function is disabled(in device parameter), screen display as below
6.6  F2 Function (no load test)
Press F2 key will begin ‘no load test’, LCD screen display as below
(  If ‘Find mark’ function is enabled(in system parameter)  ):
If ‘Search mark’ function is enabled(in device parameter), embedded software will process ‘Search mark’ function firstly, if ‘Search mark’ function is disabled, this screen won’t display.
If ‘Search mark’ function is disabled(in device parameter), screen display as below
6.7  Error test
Step 1: User should set tested meter’s parameters correctly!
(refer to 6.4.2  Set Parameters of tested meter [Menu item 1])
                     Step 2: select Output current (Ib%); select power factor(1.0L/0.5L/0.8C); 
(refer to 6.4.2  Set Parameters of tested meter [Menu item 1])
                          Step 3:  Press ‘START’ key will begin ‘Error test’, LCD screen display as below:
7. Wire connection
1. Three-phase, three-wire energy meter should be connected as below. All three-phase three-wire energy meter are connected as this picture.
2. Three-phase, four-wire energy meter should be connected as below. All three-phase four-wire energy meter are connected as this picture.
(Direct or transformer energy meter, connection diagram is same)
8.  Support & FAQs
1) When the voltage circuit is shortcut or voltage amplifier is overloaded , or the current circuit is open, or current amplifier is overloaded, testbench will automatically lower down the output until to zero and  display “ERR” ,It is recommended to lower down the output power, check the wire connection, then restart the power source output if everything is normal.
2) error test is abnormal
Please check whether the preset parameter for the tested meter and the system parameters are correct. Check whether the infrared collector works correctly
3) PTC-8000(Error calculator) doesn’t work correctly
If  PTC-8000(Error calculator) doesn’t work correctly , it can be reset by pushing the button of  “system reset” . In the process of calibrating meter, push the button of “error reset” to re-calculator Error.
4) please contact us if there is any malfunction of the equipment.
If testbench is alarm , please press ‘system reset’ to clear output firstly , after the voice of alarm is stopped, then please check current circuit and check voltage connect wire . the current circuit shouldn’t be open and the output voltage shouldn’t be shortcut.
Warning!!!
There are high voltage at Output and input of the amplifier boxes . It is dangerous to touch by hand. It is forbidden to open amplifier boxes by nonprofessional .